Flame Spectrophotometer LB-10FSP

Flame Spectrophotometer LB-10FSP

Flame Spectrophotometer LB-10FSP is a tabletop and compact flame photometer for the simultaneous measurement of elements K, Na, Li, Ca, Ba and K. Features automatic calculation of correlation coefficient and pre-selection of flame gases for reliable operation. With 8 sec of response time and up to 6 ml/min of sample uptake, has LCD touch screen display and offers direct concentration read out for easy operation.

Get Quote Download Catalog

Specifications

Measuring range K: 0.195 ppm (0.39 to 3.12) ppm
Na: 0.69 ppm (1.15 to 9.2) ppm
Li: 0.15 ppm (0.25 to 5) ppm
Ca: 3 ppm (5 to 100) ppm
Ba: 9 ppm
Sensitivity K: 0.01 ppm
Na:0.01 ppm
Li: 0.1 ppm
Ca: 2 ppm
Ba: 6 ppm
Repeatability 1% CV for 20 consecutive samples
Sample uptake < 6 ml/min
Response time < 8 sec
Packaging dimension (W×D×H) 570×530×389 mm
Gross weight 18 kg

Features

  • Offers simultaneous measurement of elements K, Na, Li, Ca, Ba and K
  • Automatic calculation of correlation coefficient for reliable operation
  • 8 sec of response time and up to 6 ml/min of sample uptake
  • High-precision unit for reproducible measurement results
  • LPG as burning gas with flameout protection and flame pre-selection
  • LCD touch screen display for convenient and reliable operation
  • Direct concentration read out for the elements

Standard Accessories

Air Compressor

Optional Accessories

Printer
Software

Applications

Flame Spectrophotometer is used for alkaline and alkaline earth elements detection across chemicals , soils, agriculture, pharmaceuticals, glass and ceramics, plant materials and water, oceanography, biological and microbiological laboratories etc.

Get Quote

Flame Spectrophotometer

Elements detected K, Na, Ca
Sensitivity K- 0.01 ppm
Na- 0.01 ppm
Ca- 2 ppm
Range K- 0 ppm ~ 100 ppm
Na- 0 ppm ~ 160 ppm
Ca- 0 ppm ~ 1000 ppm
Method detection limit K- ≤0.156 ppm
Na- ≤0.184 ppm
Ca- ≤2 ppm
Elements detected K, Na
Sensitivity K- 0.01 ppm
Na- 0.01 ppm
Range K- 0 ppm ~ 100 ppm
Na- 0 ppm ~ 160 ppm
Method detection limit K- ≤0.156 ppm
Na- ≤0.184 ppm
Measurement Range Potassium (K): 0 to 100 ppm
Sodium (Na): 0 to 160 ppm
Minimum Detection Limit Potassium (K): 0.1 ppm
Sodium (Na): 0.1 ppm
Linear Error Potassium (K): ± 0.2 ppm
Sodium (Na): ± 0.5 ppm
Display Range Potassium (K): 0 to 100 μg/ml
Sodium (Na): 0 to 160 μg/ml
Elements detected K, Na, Li
Sensitivity K- 0.01 ppm
Na- 0.01 ppm
Li- 0.01 ppm
Range K- 0 ppm ~ 100 ppm
Na- 0 ppm ~ 160 ppm
Li- 0 ppm ~ 100 ppm
Method detection limit K- ≤0.156 ppm
Na- ≤0.184 ppm
Li- 0.105 ppm
Elements detected K, Na, Ba
Sensitivity K- 0.01 ppm
Na- 0.01 ppm
Ba- 6 ppm
Range K- 0 ppm ~ 100 ppm
Na- 0 ppm ~ 160 ppm
Ba- 0 ppm ~ 3000 ppm
Method detection limit K- ≤0.156 ppm
Na- ≤0.184 ppm
Ba- ≤60.28 ppm
Elements detected K, Na, Li, Ca
Sensitivity K- 0.01 ppm
Na- 0.01 ppm
Li- 0.1 ppm
Ca- 2 ppm
Range K- 0 ppm ~ 100 ppm
Na- 0 ppm ~ 160 ppm
Li- 0 ppm ~ 100 ppm
Ca- 0 ppm ~ 1000 ppm
Method detection limit K- ≤0.156 ppm
Na- ≤0.184 ppm
Li- ≤0.105 ppm
Ca- ≤2 ppm
×

Ask Quote