Flame Spectrophotometer LB-14FSP

Flame Spectrophotometer LB-14FSP

Flame Spectrophotometer LB-14FSP is an analytical instrument with pre-selection flame sizes for elemental detection. Liquefied petroleum gas is consumed for burning of the flame to produce luminescence. It is equipped with a flame protection for safety of the operator.

Get Quote Download Catalog

Specifications

Elements detected K, Na, Li, Ca
Sensitivity K- 0.01 ppm
Na- 0.01 ppm
Li- 0.1 ppm
Ca- 2 ppm
Range K- 0 ppm ~ 100 ppm
Na- 0 ppm ~ 160 ppm
Li- 0 ppm ~ 100 ppm
Ca- 0 ppm ~ 1000 ppm
Method detection limit K- ≤0.156 ppm
Na- ≤0.184 ppm
Li- ≤0.105 ppm
Ca- ≤2 ppm
Linearity K- 0.195 ppm (0.39 ppm ~ 3.12 ppm)
Na- 0.69 ppm (1.15 ppm ~ 9.2 ppm )
Li- 0.15 ppm (0.25 ppm ~ 5 ppm)
Ca- 3 ppm (5 ppm ~ 100 ppm)
Linearity error K- ≤0.195 ppm
Na- ≤0.69 ppm
Li- ≤0.147 ppm
Ca- ≤3 ppm
Sample uptake < 6 mL/min
Repeatability 1% CV for 20 consecutive samples
Response time < 8s
Ambient temperature 10 °C ~ 35 °C
Humidity ≤ 85%
Power supply AC 110 V ±10%; 50 Hz
AC 220 V ±10%; 60 Hz
Power consumption 250 W
Dimensions 580 × 540 × 400 mm
Weight 18 kgs

Features

  • Adjustable measuring range
  • Direct concentration read out
  • Large LCD touch screen display for ease of accessibility
  • Correlation coefficient with automated calculative function (Except LB-10FSP)
  • Equipped with an air compressor for stable flow rate of air during flammation
  • USB interface (Except LB-10FSP)

Standard Accessories

Sr. No.Name of the accessoryQuantity
1.Air compressor1 unit

Applications

Used in the fields agriculture, chemistry, pharmacology, pathology, biology, forensics, etc. for analysis of particular elements in the material.

Get Quote

Flame Spectrophotometer

Elements detected K, Na, Ba
Sensitivity K- 0.01 ppm
Na- 0.01 ppm
Ba- 6 ppm
Range K- 0 ppm ~ 100 ppm
Na- 0 ppm ~ 160 ppm
Ba- 0 ppm ~ 3000 ppm
Method detection limit K- ≤0.156 ppm
Na- ≤0.184 ppm
Ba- ≤60.28 ppm
Elements detected K, Na, Li
Sensitivity K- 0.01 ppm
Na- 0.01 ppm
Li- 0.01 ppm
Range K- 0 ppm ~ 100 ppm
Na- 0 ppm ~ 160 ppm
Li- 0 ppm ~ 100 ppm
Method detection limit K- ≤0.156 ppm
Na- ≤0.184 ppm
Li- 0.105 ppm
Elements detected K, Na, Ca
Sensitivity K- 0.01 ppm
Na- 0.01 ppm
Ca- 2 ppm
Range K- 0 ppm ~ 100 ppm
Na- 0 ppm ~ 160 ppm
Ca- 0 ppm ~ 1000 ppm
Method detection limit K- ≤0.156 ppm
Na- ≤0.184 ppm
Ca- ≤2 ppm
Measurement Range Potassium (K): 0 to 100 ppm
Sodium (Na): 0 to 160 ppm
Minimum Detection Limit Potassium (K): 0.1 ppm
Sodium (Na): 0.1 ppm
Linear Error Potassium (K): ± 0.2 ppm
Sodium (Na): ± 0.5 ppm
Display Range Potassium (K): 0 to 100 μg/ml
Sodium (Na): 0 to 160 μg/ml
Elements detected K, Na
Sensitivity K- 0.01 ppm
Na- 0.01 ppm
Range K- 0 ppm ~ 100 ppm
Na- 0 ppm ~ 160 ppm
Method detection limit K- ≤0.156 ppm
Na- ≤0.184 ppm
Measuring range K: 0.195 ppm (0.39 to 3.12) ppm
Na: 0.69 ppm (1.15 to 9.2) ppm
Li: 0.15 ppm (0.25 to 5) ppm
Ca: 3 ppm (5 to 100) ppm
Ba: 9 ppm
Sensitivity K: 0.01 ppm
Na:0.01 ppm
Li: 0.1 ppm
Ca: 2 ppm
Ba: 6 ppm
Repeatability 1% CV for 20 consecutive samples
Sample uptake < 6 ml/min
×

Ask Quote

Copyright © Labotronics Scientific 2026